MRI Compatible Retention System for a Bone Conduction Device: An evaluation of required design changes to Sentio Ti™ Implant for compliance with 3 T MRI scans
Hämtar...
Publicerad
Författare
Typ
Examensarbete för masterexamen
Master's Thesis
Master's Thesis
Modellbyggare
Tidskriftstitel
ISSN
Volymtitel
Utgivare
Sammanfattning
Magnetic resonance imaging (MRI) compatibility is an important requirement for
implantable hearing devices, as increasing numbers of patients are expected to undergo
MRI examinations during their lifetime. This study evaluates the mechanical
response of the Sentio Ti™ transcutaneous bone conduction implant under magnetic
torque corresponding to a 3 T MRI environment, with the aim of assessing
whether the current design meets established performance criteria defined at 1.5 T.
A computational approach was conducted using an existing, experimentally validated
finite element model developed in LS-DYNA. The model was used to simulate
implant displacement and contact pressure on surrounding soft tissue under
worst-case magnetic torque conditions. The torque at 3 T was estimated based on
proportional scaling from 1.5 T. Parametric studies were conducted to investigate
the influence of reinforcement wire properties and silicone stiffness on implant behaviour.
In addition, alternative retention magnet concepts were explored through
a concept generation process. The results show that increasing the magnetic field
strength from 1.5 T to 3 T leads to a significant increase in mechanical response,
with displacement rising from 3.12 mm to 5.65 mm and average contact pressure
increasing by a factor of approximately 2–3. Among the investigated parameters,
reinforcement wire diameter was found to have the greatest influence on reducing
both displacement and contact pressure. However, achieving equivalent performance
to 1.5 T through structural modifications alone requires design changes that may be
impractical within current geometric constraints. Combined modifications of wire
diameter and silicone stiffness provided more feasible solutions, although they did
not fully replicate baseline pressure levels. The findings indicate that while structural
optimization can significantly improve performance, it may not be sufficient to
ensure MRI compatibility at 3 T without compromising design constraints. Modifications
to the retention magnet system, such as enabling rotational alignment or
controlled movement, are therefore identified as promising strategies. This work provides
quantitative insight into implant behaviour at higher magnetic field strengths
and supports the development of next-generation MRI-compatible bone conduction
devices.
Beskrivning
Ämne/nyckelord
MRI compatibility, bone conduction, transcutaneous implant, magnetic torque, FEM, LS-Dyna, contact pressure, implant displacement, retention magnet, Sentio Ti™