Design of smart orthosis for rehabilitation of Achilles tendon ruptures

dc.contributor.authorAnderson, Adam
dc.contributor.departmentChalmers tekniska högskola / Institutionen för elektrotekniksv
dc.contributor.examinerDean, Emmanuel
dc.contributor.supervisorBrorsson, Annelie
dc.contributor.supervisorNilsson Helander, Katarina
dc.date.accessioned2025-02-04T13:10:31Z
dc.date.available2025-02-04T13:10:31Z
dc.date.issued2025
dc.date.submitted
dc.description.abstractAbstract This thesis presents a compact system for measuring forces under the foot, designed for rehabilitation after an Achilles tendon rupture. The aim is that this system could provide patients and care providers with critical data about the recovery process, for a more personalized and effective treatment. The core component of this system is a newly developed flexible insole that senses forces under the foot. The force is measured in three dimensions (i.e., normal and shear forces) using magnetic-based sensors, placed in a grid of 73 nodes. The large area covered by the sensor, and the flexibility, are improvements over previous magnetic-based force measurement systems. The insole and additional support electronics were mounted on a standard ankle orthosis (also known as Walker). In addition, two IMUs were used to estimate the orientation of the insole. Software was also developed to process and visualize the data. The measurements from the insole are sent to a signal processing chain to calculate relevant biomechanics parameters such as the center of pressure and joint torques. The signal processing chain was implemented within ROS2, together with micro-ros for the low-level communication with hardware. ROS2 is also used for visualization purposes. The results are promising, showing that magnetic-based sensors are feasible for measuring 3D forces under the foot. The sensors display a nearly linear response to vertical pressure, although there is considerable hysteresis that introduces errors in the measurements. Future work to improve calibration, verify reliability, and improve ease of use is needed before the system can be used in a clinical setting.
dc.identifier.coursecodeEENX30
dc.identifier.urihttp://hdl.handle.net/20.500.12380/309106
dc.language.isoeng
dc.setspec.uppsokTechnology
dc.subjectKeywords: Plantar pressure, shear force, ROS2, micro-ros, Achilles’ Tendon Rupture, rehabilitation, Hall-effect.
dc.titleDesign of smart orthosis for rehabilitation of Achilles tendon ruptures
dc.type.degreeExamensarbete för masterexamensv
dc.type.degreeMaster's Thesisen
dc.type.uppsokH
local.programmeSystems, control and mechatronics (MPSYS), MSc
Ladda ner
Original bundle
Visar 1 - 1 av 1
Hämtar...
Bild (thumbnail)
Namn:
Master_s_thesis_Adam_smart_orthosis.pdf
Storlek:
2.39 MB
Format:
Adobe Portable Document Format
Beskrivning:
License bundle
Visar 1 - 1 av 1
Hämtar...
Bild (thumbnail)
Namn:
license.txt
Storlek:
2.35 KB
Format:
Item-specific license agreed upon to submission
Beskrivning: